-
2 × $233.00
-
1 × $20,326.00
-
1 × $959.00
-
1 × $1,337.00
-
1 × $955.00
-
1 × $176.00
-
1 × $205.00
-
1 × $919.00
-
1 × $9,744.00
-
1 × $134.00
-
1 × $113.00
-
1 × $3,131.00
-
1 × $163.00
-
Instek PSW-1080L125 DC Power Supply, Triple CH, Prog. Switching, 30/40/160V, 1080W, PSW-Multi Series1 × $2,175.00
-
1 × $1,337.00
-
1 × $3,091.00
-
1 × $1,177.00
-
1 × $595.00
-
1 × $997.00
-
1 × $4,416.00
-
1 × $555.00
-
1 × $959.00
-
1 × $4,306.00
-
1 × $1,407.00
-
1 × $2,404.00
-
1 × $2,162.00
-
2 × $959.00
-
1 × $5,446.00
-
1 × $4,793.00
-
1 × $955.00
-
1 × $457.00
-
1 × $1,177.00
-
1 × $1,337.00
-
1 × $17,859.00
-
1 × $2,287.00
- 5-range testing voltage 50 V/100 MO to 1000 V/4000 Mohms
- Stable & high-speed digital readings, response time at PASS/FAIL decision 0.3 second
- Drop proof onto concrete from 1m (3.28 feet)
- Bright LED luminous LCD, Test lead with bright LED lamp to illuminate near hand (Option L9788-11 or L9788-10)
- Continuity check via 200 mA testing
- Built in AC/DC voltage meter, useful for testing solar power generation systems and electric vehicles
- DC voltage measurement 4.2V (0.001V resolution) to 600V (1V resolution), 4 ranges, Accuracy: ±1.3% rdg. ±4 dgt., Input resistance: 100 kO or higher
- AC voltage measurement 420V (0.1V resolution) / 600V (1V resolution), 2 ranges, 50/60 Hz, Accuracy: ±2.3% rdg. ±8 dgt., Input resistance: 100 kO or higher, Average rectifier
- Low resistance measurement For checking the continuity of ground wiring, 10 O (0.01 O resolution) to 1000 O (1 O resolution), 3 ranges, Basic accuracy: ±3 % rdg. ±2 dgt., testing current 200 mA or more (at 6 O or less)
- Display Semi-transmissive FSTN LCD with back lighting, bar-graph indicator
- Response time Approx. 0.3 second for PASS/FAIL decision (at HIOKIs conditions)
- Other functions Indicate MO measurement value after a lapse of one minute, Live circuit indicator, Automatic electric discharge, Automatic DC/AC detection, Comparator, Drop proof, Auto power save
- Power supply LR6 (AA) alkaline batteries × 4 Continuous use: 20 hours (at HIOKI’s conditions) Number of measurements: 1000 times (at 5 s ON, 25 s OFF cycle, insulation measure of lower limit resistance value to maintain nominal output voltage)
- Dimensions, mass 159 mm (6.26 in) W × 177 mm (6.97 in) H× 53 mm (2.09 in) D, 640 g (22.6 oz) (including batteries, excluding test leads)
>
Each Unit Includes
- Test Lead
- Neck strap
- Instruction Manual
- 4 LR6 (AA) alkaline batteries











Reviews
There are no reviews yet.